VITEK®MS
MALDI-TOF Microbial Identification System :
NEW : Expanded knowledge base !
The VITEK® MS IND system is the ultimate ID technology by the inventor of automated phenotypic identification.
- 86 new fungi : the most reliable fungi identifications using maldi-tof mass technology
- 322 new bacteria : the most robust aerobic, anaerobic microaerophic identifications.
- Ultra fast identification with efficient workflow
Get results in minutes !
- Pick colony
- Direct deposit on the slide
- Add ready-to-use CCHA Matrix (Formic Acid – Yeast)
- Load into VITEK® MS
- Species represented by population: specific species’ characteristics can be revealed
- Average of 18 isolates per species from different fields (clinical, food and pharma) and geographical origins (Asia, America, Europe)
- Average of 40 spectra per species: several culture media and instruments
- Ultra fast time-to-result: minutes for a reliable result
- High throughput: identify up to 192 isolates per run
- Less hands-on time: simpler workflow
- Easy-to-use: no gram staining and microscopic observation is required
- Fast and flexible method: multiple users can prepare samples simultaneously to save time
- High performing identification
- Informative and easy-to-read results
- Traceability: slides are barcoded and software is easy-to-use
Mass spectrometry for microbial identification: the MALDI-TOF technology
he VITEK® MS system relies on mass spectrometry, an analytical technique which determines the elemental composition of a sample. It is the latest technology for microbial identification.
MALDI-TOF: Matrix Assisted Laser Desorption Ionization – Time of Flight analyzer
To perform identification, the sample to be analyzed is mixed with another compound : the matrix. When the mixture is applied to a metal plate and irradiated with a laser, the matrix vaporizes and the sample gains an electrical charge, whose mass-to-charge ratio is measured.
To identify microorganisms, the VITEK® MS:
- Ionizes the sample thanks to matrix and laser reaction
- Uses an electric field to accelerate the ions through the same potential
- Measures the time they take to reach the detector – Time of Flight
- Spectrum is sent and compared to a database
- 337 nm nitrogen laser, fixed focus
- Maximum pulse rate - 50 Hz (50 laser shots per second)
- Near normal (on-axis) incidence of the laser beam to the sample
- Laser power and laser aim under software control
- Linear flight tube of 1.2 m drift length
- Vacuum maintained by two turbomolecular pumps (nominal 250 l/s) with rotary backing
- Beam blanking to deflect unwanted high intensity signals e.g. matrix ions
- Size (w h d) - 0.7 m x 1.92 m x 0.85m, minimum distance to wall at back is 100 mm
- Weight - 330 kg excluding data system
- Electrical
- A ‘clean’, stable and continuous mains supply is required for reliable operation
- Temperature - ambient 18° to 26° C
- Relative humidity - less than 70% non-condensing
- Vibration free, firm, level floor, at least 330 kg supported at four points
- 1 to 500 kDa